MINIMIZING RADIATION
EXPOSURE
Know where the radiation is & where it is
coming from.
| Very
little radiation comes directly from the x-ray source – outside of the
collimated primary beam – since the x-ray tube is surrounded by lead
shielding.
Therefore, most radiation you will receive will be
x-rays scattered from the patient
(unless you place yourself within the direct primary
beam from the x-ray tube).
|
|
-
Scattered x-rays from the patient are most intense on the radiation
entrance side of the patient.
-
Very important
The amount of scatter
produced is proportional to the collimated area.
Reduce the collimation to reduce the scatter!
Keep
your hands out of the primary beam
-
If your hands might intrude into the primary beam, then wear lead
gloves.
-
The primary beam in fluoroscopy may be larger than what you see on the
image monitor.
-
For example, with GE fluoros the radiation field fills
the square around the circle that is imaged – there is no iris
(circular)
collimation. Also, if a fluoro
machine is malfunctioning, and has not been tested for proper collimation, the
actual
radiation field may be MUCH larger than the imaged field.
I have seen radiation fields as much as 8
times the size of the imaged field – which were then quickly repaired.
Distance
-
Radiation intensity decreases as the distance squared. If
the dose rate at 1 meter is 16 mrem/min, then the dose
rate at 4 meters is only 1 mrem/min.
-
If you must be in the patient room when the x-rays are
activated, step away from the patient as much as possible.
Time
-
The dose increases as the amount of time you are exposed to the radiation
increases.
-
Minimize the amount of time that you are in the patient
room when the x-rays are activated. Minimize the amount
of time the patient is exposed to x-rays.
Shielding
-
Your dose is reduced by placing an attenuating material
between you and the source of x-rays.
-
Shielding can involve placing another person between you
and the source of scattered radiation.
-
Personal Shielding Items:
- Lead Apron
- Thyroid Shield
- Lead Gloves
- Leaded Glasses
The above normally have 0.5 mm lead equivalent
which allows approximately a 2% transmittance
of incident radiation.
- Equipment and Room Shielding:
- Lead shielding drapes on x-ray equipment or
- mounted on arm to ceiling
- Portable radiation shields (on wheels)
- Leaded glass windows
- Leaded walls